eChip® – The Oximetry Module

eChip® – The Oximetry Module

  • Accuracy proven by Bickler Laboratory of University of California San Francisco.
  • Small Size allows it to fit into small handheld devices.
  • Low Power Consumption which facilitating long battery life in handheld devices.
  • Signal Index (Perfusion level) is providing a proportional to the perfusion level of the patient
  • Plethysmogram is automatically scaled and offset in real time to fit into graphical display. A real-time pulse bar output is provided to indicate the strength of pulse.
  • 3 response modes selection; Sensitive. Stardard & Stable.
  • UART Communication which provided a flexible and straightforward data output making it implement in the host system easily
  • A real-time warning and error reporting system, allowing the host to monitor whether eChip® is working properly. Messages and warnings are provided for abnormal operation.

Description

  • Accuracy proven by Bickler Laboratory of University of California San Francisco.
  • Small Size allows it to fit into small handheld devices.
  • Low Power Consumption which facilitating long battery life in handheld devices.
  • Signal Index (Perfusion level) is providing a proportional to the perfusion level of the patient
  • Plethysmogram is automatically scaled and offset in real time to fit into graphical display. A real-time pulse bar output is provided to indicate the strength of pulse.
  • 3 response modes selection; Sensitive. Stardard & Stable.
  • UART Communication which provided a flexible and straightforward data output making it implement in the host system easily
  • A real-time warning and error reporting system, allowing the host to monitor whether eChip® is working properly. Messages and warnings are provided for abnormal operation.
Parameter Range Unit
Min. Typ. Max.
SpO2 0 – 100 %
SpO2 accuracy70 – 99%

<70%

–

–

±2

Monotonic

–

–

%
SpO2 resolution – 1 – %
Pulse rate range 30 – 249 BPM
Pulse rate resolution ±3 BPM
Perfusion range 0.05 – 20 %
SpO2 averagingStable

Normal

Sensitive

–

–

–

16

12

6

–

–

–

Pulse
Motion artifact – Not Specified – /
Signal index – NormalReduced

Poor

–
Pulse wave 0 – 255 LSB
Pulse bar 0 – 15 LSB
Other specifications
Parameter Range   Unit
Min. Typ. Max.
Operating Temperature: -20 – +60 °C
Storage Temperature -30 – +70 °C
Operating humidity (Non Condensing) 0 – 90 %
Storage humidity (Non Condensing) 0 – 90 %
Dimensions – 18 x 31 x 6 – mm
Weight – 2 – g